Control Flow Strategy in a Receiver Coil for Nuclear Magnetic Resonance for Imaging

A mathematical discussion is introduced to describe the receiver coil characterizing the NMRI system starting from a general shape of the conductor. A set of different inductance calculations have been introduced varying the shape of the conductor. The inductance calculation led to a general express...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autor principal: Barbaraci, Gabriele
Formato: Capítulo de libro
Lenguaje:inglés
Publicado: IntechOpen 2020
Materias:
Acceso en línea:https://www.intechopen.com/chapters/71443
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:A mathematical discussion is introduced to describe the receiver coil characterizing the NMRI system starting from a general shape of the conductor. A set of different inductance calculations have been introduced varying the shape of the conductor. The inductance calculation led to a general expression of the magnetic field of a single coil characterized by a rectangular shape. A dynamic model of the receiver coil has been developed to represent the natural frequencies that characterize the operational bandwidth. A nonstationary control strategy is implemented to make a real-time changing of the operational bandwidth. The frequency response of the coil generates the necessary conditional expression in order to let the peak of resonance move to a desired value of frequency.
ISBN:978-1-83880-419-0
DOI:10.5772/intechopen.88561