Special Cuvettes for Spectrofluorimeters

Special cuvettes that dramatically increase the fluorescence intensity of diluted solutions and suspensions of isolated biomembranes and living cells are described. (1) Mirror cuvettes have a reflecting aluminum or silver layer, putting on their three external sides, having a slot-hole window at fro...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autor principal: Vekshin, Nikolai
Formato: Capítulo de libro
Lenguaje:inglés
Publicado: IntechOpen 2020
Materias:
Acceso en línea:https://www.intechopen.com/chapters/72344
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Descripción
Sumario:Special cuvettes that dramatically increase the fluorescence intensity of diluted solutions and suspensions of isolated biomembranes and living cells are described. (1) Mirror cuvettes have a reflecting aluminum or silver layer, putting on their three external sides, having a slot-hole window at frontal side center. (2) Mirror microcuvettes have two reflecting sides. (3) Total internal reflection cuvettes have lateral reflecting side and two triangular prisms, attached outside; the cuvette works effectively with a parallel excitation light beam. (4) In cuvette with diagonal quartz plate, very small quantity of a sample is attached in the form of a thin layer. (5) Mirror cuvette with diagonal plate. All described cuvettes can be applied for measurements of spectra excitation and emission, excited-state lifetime, and polarization degree. Also, they can be used for photo-bleaching of biological chromophores.
ISBN:978-1-83968-039-7
DOI:10.5772/intechopen.92720